What Is Measurement System Analysis (MSA)?

Definition

Measurement System Analysis (MSA) is a set of studies that show how much of the variation in your measurements comes from the measurement system itself, so you know whether the data can be trusted.

MSA in Simple Terms

Every measurement includes some error from the gauge, the operator, the method and the environment. If that error is large compared with the tolerance, good parts can be rejected and bad parts accepted. MSA measures that error. It is one of the AIAG core tools and is required during PPAP for the gauges named in the control plan.

Key Elements

StudyWhat it checks
BiasDifference between the average reading and a reference value
LinearityWhether bias changes across the measuring range
StabilityWhether the system drifts over time
Gauge R&RRepeatability (gauge) and reproducibility (operators)
Attribute agreementConsistency of go/no-go or visual decisions

In the AIAG MSA manual, a Gauge R&R under 10 percent is generally acceptable, 10 to 30 percent may be acceptable depending on the application, and over 30 percent is unacceptable. The number of distinct categories should be 5 or more.

Example

Three operators measure 10 shafts twice each with a micrometer. The study gives a GRR of 24 percent of tolerance, mostly from reproducibility. The team finds operators measure at different points along the shaft, adds a fixture and a clear method to the work instruction, and a repeat study gives 8 percent.

Common Mistakes

  • Choosing parts that do not cover the real process range.
  • Running the study only once, at PPAP, and never again.
  • Confusing MSA with calibration, which checks accuracy against a standard but not the whole system.

Gauge R&R and MSA records can be kept with calibration data in OperaHive calibration management software.

FAQ

Frequently Asked Questions

Still unsure? Ask our team.

What is an acceptable MSA result?

Under the AIAG MSA manual, a GRR below 10 percent is generally acceptable, 10 to 30 percent may be acceptable depending on the application and cost, and above 30 percent needs improvement.

What is the difference between MSA and Gauge R&R?

Gauge R&R is one type of MSA study. MSA also covers bias, linearity, stability and attribute agreement.

What is MSA in quality?

MSA, or measurement system analysis, is a set of studies that show how much of the variation you see in measurements comes from the measurement system rather than the parts.

What does MSA stand for?

MSA stands for measurement system analysis. The AIAG MSA manual is the main reference in the automotive industry.

What are the types of MSA studies?

Common studies are bias, linearity, stability, gauge repeatability and reproducibility (gauge R&R) for variable data, and attribute agreement analysis for pass or fail checks.

What is an acceptable gauge R&R result?

Under AIAG guidance, under 10 percent is acceptable, 10 to 30 percent may be acceptable depending on the application, and over 30 percent is not acceptable. The number of distinct categories should be 5 or more.

How do you perform a gauge R&R study?

A typical study uses 10 parts, 3 appraisers and 2 or 3 trials. Each appraiser measures each part in random order, then the results are analysed by ANOVA or the average and range method.

Is MSA required by IATF 16949?

Yes. IATF 16949 requires statistical studies to analyse variation in each type of measurement system named in the control plan, using methods accepted by the customer.

When should MSA be repeated?

Repeat it when a gauge is new, repaired or modified, when the process or product changes, when operators change significantly or at intervals set by the plant.

What is the difference between MSA and calibration?

Calibration checks a gauge against a known reference. MSA checks how the whole system, including gauge, people, method and environment, performs on real parts.