Calibration in Simple Terms
Every inspection decision depends on the gauge being right. Calibration proves that a caliper, micrometer, torque wrench or CMM reads within its allowed error. The reference standards must be traceable to national or international standards, usually through an accredited lab working to ISO/IEC 17025.
ISO 9001, IATF 16949 and AS9100 all require measuring equipment to be calibrated or verified at set intervals, identified with its status and protected from adjustments that would make results invalid.
Key Elements
- A unique ID for every gauge and a master list of all gauges.
- A calibration interval based on use, risk and history.
- A calibration record showing as-found and as-left readings, the standard used and its traceability.
- A status label or system flag so expired gauges are not used.
- A procedure for gauges found out of tolerance.
Example
A bore gauge comes back from calibration reading 0.015 mm low. Because it was used to accept parts since the last calibration, the quality team reviews the parts measured in that period, finds two lots close to the lower limit, and re-inspects them with a good gauge before deciding whether the customer must be told.
Common Mistakes
- Recording only pass or fail, with no as-found readings.
- Not assessing product measured with a gauge later found out of tolerance.
- Forgetting employee-owned tools and test software.
- Fixed intervals that are never reviewed against history.
Calibration covers accuracy, while MSA checks the whole measurement system, including bias. OperaHive calibration management software tracks schedules, due dates and certificates and helps investigate failed gauges.